| ACWL/IS - Overview |

|
 |

|
Autoclavable combination pH electrode with inclined sidearm
Best for: Benchtop fermenters
ACWL/IS/A=120 - Screw cap, 120 mm
'A' lengths 150, 200, 250, 300 and 500 also available
Benchtop fermenters are usually autoclaved at 120-135 o C prior to use. If pH or ORP monitoring is required then the electrodes must be calibrated and then autoclaved with the complete system. This puts considerable demands on pH (and ORP) electrodes.
Thermo Electron has developed a unique formulation of pH glass for this purpose. This stands up to the rigours of autoclaving much better than conventional pH glass with a significantly lower change in calibration values. The silver/silver chloride reference system has to be of more robust construction to prevent degradation of the silver wire. A viscous formulation of reference electrolyte is used to reduce loss. Extra seals are used in the cap construction that is fully steam resistant.
Electrodes are supplied complete with a blanking cap. |
 |
|
 |
 |
| Sales Contact |
Tel: +44 (0)1337 828871
Fax: +44 (0)1337 828972
e-mail: sales.water.uk@thermo.com
or technical@thermorussell.com |
| Spec Sheets Downloads |
|
ACWL
|
| Specifications |
| Temp range (°C) |
0-130 |
| Reference Junction |
Ceramic frit |
| Viscous Reference Electrolyte |
S00R/250 |
| Dimensions |
12mm diameter and 'A' length (below sidearm) user selected |
| Cable Length |
NA |
| Connector type |
S7 |
|
|